ACOUSTOCAM™ SPECS
General
Dimensions
(W x H x D)
Weight
Camera Cable
Battery
Power Requirements
Weight
Camera: 7.8” x 3.5” x 3.0” (19.8 cm x 8.8 cm x 7.6 cm)
Controller: 12.5” x 8.0” x 3.5” (31.7 cm x 20.3 cm x 8.8 cm)
Camera: 1.3 lbs. (0.6 kg)
Controller: 6.0 pounds (3.2 kg)
8.0’ (2.4 m) length; customized lengths to 20’ (6.1 m)
1 gigabyte RAM; 120 gigabyte hard drive
Rechargeable 14.4 volt lithium-ion; 90 minute battery life;
built-in charger
110/240 VAC; 50/60Hz
Intel Atom 1.86GHz processor; Microsoft Windows XP Pro
C-Scan: pulse-echo
A-Scan: pulse-echo, dual element
Inputs/Outputs
System Hardware & Software
Data Storage
Display
High brightness TFT color LCD; SVGA; 800 W x 600 H; 6.75” W x 5.25” H; 8.4” diagonal (17cm x 13cm; 21.6cm)
Sunlight readable
Display Modes
C-Scan; A-Scan; combined C/A-Scans
Normal; Fullscreen
Software
AcoustoVision™
Touchscreen GUI
Ethernet
100 BaseT
(1) RJ45, F jack
Wirless (Optional)
802.11 a/b/g (supports remote NDI system)
(1) 4.25” antenna
VGA Controller
VGA 800 W x 600 H
(1) 15-pin HD
D-sub, F
USB
USB 2.0
(1) USB jack
C-Scan
Acoustic Lenses
Three element F/1 lens set
Acoustic Field of View
1.0” x 1.0” (2.5 cm x 2.5 cm) instantaneous, 20” x 16” maximum scan
Imaging ASIC Array
Patented Digital Acoustic Video™ Version 6.0
Array Elements
120 x 120 (14,400 pixels)
Pixel Size
100 microns
C-Scan Video Rate
30 frames per second
X/Y Spatial Resolution
600 microns (0.6 mm) @ 5 MHz
Curvature
4” to flat using pipe trolley
Thickness Range
0” to 3.0” (0 cm to 7.5 cm)
Dynamic Range
60 dB instantaneous
Depth Resolution
10 ns
Image Acquisition Range
Variable delay from 0 µs to 250 µs
Variable detection window from 0 ns to end of range in 10 ns increments
Match Gates
C-Scan gate start/width can match to A-Scan gate start/width
X,Y Defect Sizing
User calibrated grid or point-to-point manual sizing
Angle of Inspection
0 degree
Image Capture
C-Scan; A-Scan; combined C/A-Scans
Image Capture Type
PNG; on-board storage up to 100,000 image capture files
Video Capture Type
MPEG-3; 60 s max. record time/file; on-board storage up to 50,000 files
A-Scan
Gain
0 to 60 dB
Maximum Input Signal
5 Vp-p
Receiver Input Impedance
10 MΩ ± 5%
Receiver Bandwidth
0.1 to 20 MHz @ -3 dB
A-Scan Rectification
Full; half; none
Receiver Delay
0 to 0.2 µs
Digital Filter Settings
Yes
Reject
Variable from 0% to 100% FSH
Time Corrected Gain
Up to 5 reflected peaks can be used for calibrating TCG
Amplitude Measurement
0% to 110% full screen height with 0.125% resolution
Measurement Rate
30 Hz (equivalent to PRF)
A-Scan Gate Start
Variable over entire displayed range
A-Scan Gate Width
Variable from gate start to end of displayed range
A-Scan Gate Height
Variable from 1% to 100% FSH
Gate Type
Peak; Flank; Zero Crossing; Back Wall
Alarms
Alarm feature for 2nd reference cursor
Acoustic Velocities
Pre-installed, editable library of 40 commonly inspected materials plus user added materials
Pulser
Pulses per Burst
Up to 20 square wave pulses
Pulser Voltage
0 V to 1000 V
Pulse Repetition Frequency
250 Hz
Pulser Rise Time
< 15 ns
Bandwidth
Adjustable from 0.5 MHz to 7.5 MHz
External Transducer
Dual-element, 5MHz
2-Dimensional Scanning & Stitching
Encoder
1.25” (31.75 mm) diameter wheels; 20 steps/mm resolution
Scanning Software
AcoustoVision™
Scan Imaging Features
Real-time display; image auto-correction;
on-board storage up to 600,000 scans
Scan Area
12.0” x 12.0” (50 cm x 50 cm) per image
Scan Speed
6”/second
Certifications & Environmental Qualifications
ASTM E317-06a
Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Testing Instruments and Systems
ASTM 1324-11
Standard Guide for Measuring Some Electronic Characteristics of Ultrasonic Testing Instruments
Operating Temperature
Normal Operating Temperature: 32 °F to 122 °F (0 °C to 50 °C)
Camera Head Contact Temperature: Up to 200 °F (93 °C)